No.ComplaintReasonsRemedy
1Machinedoesnot work;powerindicator doesnotlightupat powerswitchon position.1.NoACsupply 2.MCBtripped 3.Mainscordfailure 4.Powerswitchfailure1.RestoreACsupply 2.Checkforanyshortcircuitand resettheMCB. 3.Replacemainscord. 4.Replacetheswitch.
2Vacuumpumpdoes notworkwhenlidis closedanddisplay READYMODEonly1.Reedswitchdoesnot operate. 2.Defectivereedswitch.1.Adjustthereedswitchclamp suitablytooperatetheswitch whenlidisclosed. 2.Replacethereedswitch.
3Vacuumpumpdoes notworkwhenlidis closedanddisplay showingvacuumlevel.1.Defectivecontactor (coilopenorcontactburned out). 2.Defectiverelaypcba. 3.Defectivemotor.1.Replacethecontactor. 2.Replacetherelayboard. 3.Replaceorrectifythemotor.
4Theliddoesnot remainclosedeven aftervacuumpump starts1.Lidgasketdoesnottouch thechamberfaceuniformly.1.Pressthelidwithboththehands atthecorneroflid. 2.Checkthegasketforanydamage andreplaceifnecessary.
4InsufficientVacuum Reading1.Lowvacuumsetting 2.Solenoidvalveleaks: atmosphericairentersinto thechamberthroughthe solenoidvalvedueto, a)Defectivesolenoidvalve. b)Defectiverelay. 3.Lowoillevelinthe vacuumpump. 4.Moistercontaminatedoil. 5.Vacuumleaksthroughlid gasket/boltholesonthe lid/pipejoints/bushesat cableentry. 6.Wearandtearofvanesof vacuumpump. 7.Defectivetransducer Board1.Increasethevacuumsetting. 2.a)Replacethesolenoidvalve. b)Replacerelayboard. 3.Keeptherequiredoilqtyinthe vacuumpump. 4.Draintheoilcompletelyand refill. 5.Identifythelocationofleakage andreplacethedefective componentsorrectifytheproblem bytighteningthejoints. 6.Overhaul/replacethepump. 7.Replacethedefective.
5Noflushingoperation. Sealingoperation startsaftervacuum time.1.Flushpressuresettingsare notproper.1.Setflushpressureproperly.Set Flushlevelflush1<Vacuumlevel
6Noflushingoperation. Machinehaltsafter evacuation.Display showsgasflush.1.Defectivesolenoidvalve. 2.Verylowpressureingas cylinder. 3.Gascylindervalveclosed. 4.DefectiveRelay.1.Replacesolenoidvalve. 2.Replace/refillthegascylinder. 3. Openthegascylinder valve. 4.Replacetherelayboard/Relay.
7Machineskipsflushing andsealingoperation doesnottakesplace.1.Flushpressuresettingis high.1.Setflushpressureproperly. Flushingpressureshouldbelower thanthevacuumpressuresetting.
8Sealingheadoperate butsealingdoesnot takesplace1.Lowsealingtimesetting. 2.heatingelementfailure. 3.Defective sealingtransformer. 4.Flushsettingishigh.1.Increasesealingtime. 2.Replaceheatingelement. 3.Replacedefectivesealing transformer. 4.Usecompressorforoperating jawforflushsettingabove300.
9Machinehaltsafter sealing1.Defectiverelay. 2.Defectiveairinletsolenoid valve.1.Replacetherelayboard. 2.Replacesolenoidvalve.
10Pouchexpandsduring vacuumtime1.Lowgapbetweensealing headandpadholder assembly.1.Increasethegapbyadjustingthe padholderassembly.
11Nonuniformsealing1.Highgapbetweensealing headandpadholder assembly. 2.Sealingheadandpad holdernotalignproperly. 3.BurntTeflonfabric.1.Reducethegap. 2.Alignsealingheadandpad holder. 3.ReplacetheTeflonfabric.
12Machinesworksfine, butblank/garbage display1.Defectivedisplay. 2.FRCcableofdisplayorits connectordefective.1.Replacedisplayboard. 2.Replacecableandconnector.
13TheMicrocontroller doesnotrespondingto pushbuttonpressing.1.Defectiveswitches. 2.DefectiveFRCCableof DisplayPCB1.Replacedisplayboard. 2.Replacecableandconnector.
14Systemhangsduring operation1.Temporarybuginprogram duetonoise. 2.Defectivemicrocontroller board.1.Switchoffandrestartthe machine. 2.ReplaceMicrocontrollerboard.
15Displaydoesnot changefromfreshair toready1.Misalignmentofreed switch.1.Correctthepositionofreed switch.
16OverheatingofHeating element/Transformer1.Shortcircuitwithbodyof machine. 1.DefectiveRelay/IRF540of heaterinRelayPCB1.Makeproperwireconnections. 1.ReplaceRelayPCB.
17Machinetripswhile pumpstartsworkingor abnormalsoundof pump.1.Windingshortwithbody. 2.Defectivecentrifugal switch. 3.Defectivelovejaw couplingbush.1.Rewindorreplacethepump. 2.Replacetheswitchorpump. 3.Replacethebushorpump.
18Displayshowsfreshair always1.Misalignmentofmagnetic reedswitch. 2.Magneticreedswitch shorted.1.Correcttheposition. 2.Replacetheswitch.
19Impropersealing1.Misalignmentofsealing bed.1.Correcttheposition.
20Liddoesn’topenafter completingthe operation.1.Defectiverelayintherelay board. 2.Defectivesolenoidvalve.1.Replaceit. 2.Replaceit.