No.ComplaintReasonsRemedy
1Machinedoesnot work.1.NoACsupply 2.MCBtripped 3.Mainscordfailure 4.Powerswitchfailure1.RestoreACsupply 2.Checkforanyshortcircuitand resettheMCB. 3.Replacemainscord. 4.ReplaceDisplaypcb.
2Vacuumpumpdoes notworkwhenlidis closedanddisplay READYonly1.Reedswitchdoesnot operate. 2.Defectivereedswitch.1.Adjustthereedswitchclamp suitablytooperatetheswitch whenlidisclosed. 2.Replacethereedswitch.
3Vacuumpumpdoes notworkwhenlidis closedanddisplay showingvacuumlevel.1.Defectivecontactor (coilopenorcontactburned out). 2.Defectiverelaypcba. 3.Defectivevacuumpump.1.Replacethecontactor. 2.Replacetherelayboard. 3.Replaceorrectifythedefective.
4Theliddoesnot remainclosedeven aftervacuumpump starts1.Lidgasketdoesnottouch thechamberfaceuniformly.1.Pressthelidwithboththehands atthecorneroflid. 2.Checkthegasketforanydamage andreplaceifnecessary.
4InsufficientVacuum Reading1.Lowvacuumsetting 2.Solenoidvalveleaks: atmosphericairentersinto thechamberthroughthe solenoidvalvedueto, a)Defectivesolenoidvalve. b)Defectiverelay. 3.Lowoillevelinthe vacuumpump. 4.Moistercontaminatedoil. 5.Vacuumleaksthroughlid gasket/boltholesonthe lid/pipejoints/bushesat cableentry. 6.Wearandtearofvanesof vacuumpump. 7.Defectivemicrocontroller pcb1.Increasethevacuumsetting. 2.a)Replacethesolenoidvalve. b)Replacerelayboard. 3.Keeptherequiredoilqtyinthe vacuumpump. 4.Draintheoilcompletelyand refill. 5.Identifythelocationofleakage andreplacethedefective componentsorrectifytheproblem bytighteningthejoints. 6.Overhaul/replacethepump. 7.Replacethedefective.
5Noflushingoperation. Sealingoperation startsaftervacuum time.1.Flushpressuresettingsare notproper.1.Setflushpressureproperly.Set Flushlevel<Vacuumlevel
6Noflushingoperation. Machinehaltsafter evacuation.Display showsgasflush.1.Defectivesolenoidvalve. 2.Verylowpressureingas cylinder. 3.Gascylindervalveclosed. 4.DefectiveRelay.1.Replacesolenoidvalve. 2.Replace/refillthegascylinder. 3.Openthegascylindervalve. 4.Replacetherelayboard/Relay.
7Sealingheadoperate butsealingdoesnot takesplace1.Lowsealingtimesetting. 2.heatingelementfailure. 3.Defective sealingtransformer. 4.Flushsettingishigh.1.Increasesealingtime. 2.Replaceheatingelement. 3.Replacedefectivesealing transformer. 4.Usecompressorforoperating jawforflushsettingabove300.
8Machinehaltsafter sealing1.Defectiverelay. 2.Defectiveairinletsolenoid valve.1.Replacetherelayboard. 2.Replacesolenoidvalve.
9Pouchexpandsduring vacuumtime1.Lowgapbetweensealing headandpadholder assembly.1.Increasethegapbyadjustingthe padholderassembly.
10Nonuniformsealing1.Highgapbetweensealing headandpadholder assembly. 2.Sealingheadandpad holdernotalignproperly. 3.BurntTeflonfabric.1.Reducethegap. 2.Alignsealingheadandpad holder. 3.ReplacetheTeflonfabric.
11Machinesworksfine, butblank/garbage display1.Defectivedisplay. 2.FRCcableofdisplayorits connectordefective.1.Replacedisplayboard. 2.Replacecableandconnector.
12TheMicrocontroller doesnotrespondingto pushbuttonpressing.1.Defectiveswitches. 2.DefectiveFRCCableof DisplayPCB1.Replacedisplayboard. 2.Replacecableandconnector.
13Systemhangsduring operation1.Temporarybuginprogram duetonoise. 2.Defectivemicrocontroller board.1.Switchoffandrestartthe machine. 2.ReplaceMicrocontrollerboard.
14Displaydoesnot changefromfreshair toready1.Misalignmentofreed switch.1.Correctthepositionofreed switch.
15OverheatingofHeating element/Transformer1.Shortcircuitwithbodyof machine. 1.DefectiveRelaypcb.1.Makeproperwireconnections. 1.ReplaceRelayPCB.
16Machinetripswhile pumpstartsworkingor abnormalsoundof pump.1.Windingshortwithbody. 2.Defectivecentrifugal switch. 3.Defectivelovejaw couplingbush.1.Rewindorreplacethepump. 2.Replacetheswitchorpump. 3.Replacethebushorpump.
17Displayshowsfreshair always1.Misalignmentofmagnetic reedswitch. 2.Magneticreedswitch shorted.1.Correcttheposition. 2.Replacetheswitch.
18Impropersealing1.Misalignmentofsealing bed.1.Correcttheposition.
19Liddoesn’topenafter completingthe operation.1.Defectiverelayintherelay board. 2.Defectivesolenoidvalve.1.Replaceit. 2.Replaceit.