No.ComplaintReasonsRemedy
1Machinedoesnot work;Phasesequence LEDlightup1.Absenceofonephase. 2.Mainscordfailure. 3.MCBfailure. 4.Phasesarenotin sequence. 5.Imbalancebetween phases.1.RestoreACsupply. 2.Replacemainscord. 3.ReplacetheMCB. 4.Interchangeanytwophase lines.
2Vacuumpumpdoes notworkwhenlidis closedandvacuum levelisalways displayed1.Defectivecontactor (coilopenorcontactburned out). 2.Defectiverelaypcb. 3.Defectivevacuumpump.1.Replacethecontactor. 2.Replacetherelayboard. 5.Replaceorrectifythevacuum pump.
3Theliddoesnot remainclosedeven aftervacuumpump starts1.Lidgasketdoesnottouch thechamberfaceuniformly. 2.Defectivelidgasketa.Pressthelid. b.ReplacetheDefective.
4InsufficientVacuum Reading1.Lowvacuumsetting 2.Solenoidvalveleaks: atmosphericairentersinto thechamberthroughthe solenoidvalvedueto, a)Defectivesolenoidvalve. b)Defectiverelay. 3.Lowoillevelinthe vacuumpump. 4.Moistercontaminatedoil. 5.Vacuumleaksthroughlid gasket/boltholesonthe lid/pipejoints/bushesat cableentry. 6.Wearandtearofvanesof vacuumpump. 7.DefectiveMCPCBA1.Increasethevacuumsetting. 2.a)Replacethesolenoidvalve. b)Replacerelayboard. 3.Keeptherequiredoilqtyinthe vacuumpump. 4.Draintheoilcompletelyand refill. 5.Identifythelocationofleakage andreplacethedefective componentsorrectifytheproblem bytighteningthejoints. 6.Overhaul/replacethepump. 7.Replacethedefective.
5Sealingheadoperate butsealingdoesnot takesplace1.Lowsealingtimesetting. 2.heatingelementfailure. 3.Defective sealingtransformers.1.Increasesealingtime. 2.Replaceheatingelement. 3.Replacedefectivesealing transformers.
6Machinehaltsafter sealing1.Defectiverelaypcb. 2.Defectiveairinletsolenoid valve.1.Replacetherelayboard. 2.Replacesolenoidvalve.
7Pouchexpandsduring vacuumtime1.Lowgapbetweensealing headandpadholder assembly.Increasethegapbyadjustingthe padholderassembly.
8Nonuniformsealing1.Highgapbetweensealing headandpadholder assembly. 2.Sealingheadandpad holdernotalignproperly. 3.BurntTeflonfabric.1.Reducethegap. 2.Alignsealingheadandpad holder. 3.ReplacetheTeflonfabric.
9Machinesworksfine, butblank/garbage display1.Defectivedisplay. 2.FRCcableofdisplayorits connectordefective.1.Replacedisplayboard. 2.Replacecableandconnector.
10Settingisnotpossible. TheMicrocontroller doesnotrespondingto pushbuttonpressing.1.Defectiveswitches. 2.DefectiveFRCCableof DisplayPCB1.Replacedisplayboard. 2.Replacecableandconnector.
11Systemhangsduring operation1.Temporarybuginprogram duetonoise. 2.Defectivemicrocontroller board.1.Switchoffandrestartthe machine. 2.ReplaceMicrocontrollerboard.
12OverheatingofHeating element/Transformer1.Shortcircuitwithbodyof machine. 2.DefectiveRelay PCB/Contactor1.Makeproperwireconnections. 2.Replacethedefective.
13Machinetripswhile pumpstartsworkingor abnormalsoundof pump.1.Windingshortwithbody. 2.Defectivecentrifugal switch. 3.Defectivelovejaw couplingbush.1.Rewindorreplacethepump. 2.Replacetheswitchorpump. 3.Replacethebushorpump.
14Liddoesn’topenafter completingthe operation.1.Defectiverelayintherelay board. 2.Solenoidvalvecoilburned out. 3.Defectiveshockabsorber1.Replaceit. 2.Replacethesolenoidvalve. 3.Replaceit.
15Displayshowsfreshair always1.Defectivereedswitch.1.Replacetheswitch.