| 1 | Machine does not work; power switch does not light up. | 1.No AC supply. 2.Mains cord failure. 3.MCB failure. 4.Power switch failure. | 1.Restore AC supply. 2.Replace mains cord. 3.Replace the MCB. 4.Replace it. |
| 2 | Vacuum pump does not work when lid is closed and vacuum level is always displayed | 1.Defective contactor 2.Defective Relay board. 3.Defective motor. | 1.Replace the contactor. 2.Replace the relay board. 3.Replace or rectify the motor. |
| 3 | The lid does not remain closed even after vacuum pump starts | 1.Lid gasket does not touch the chamber face uniformly. | a.Press the lid. b.Check the height of swing arm. If the height is optimum, the gasket will touche the chamber uniformly. |
| 4 | Insufficient Vacuum Reading | 1.Low vacuum setting 2.Solenoid valve leaks: atmospheric air enters in to the chamber through the solenoid valve due to, a)Defective solenoid valve. b)Defective relay. 3.Low oil level in the vacuum pump. 4.Moister contaminated oil. | 1.Increase the vacuum setting. 2. a)Replace the solenoid valve. b)Replace relay board. 3.Keep the required oil qty in the vacuum pump. 4.Drain the oil completely and |
| | | |
| 5 | Sealing head operate but sealing does not takes place | | |
| 6 | Machine halts after sealing | | |
| 7 | Pouch expands during vacuum time | | |
| 8 | Nonuniform sealing | | |
| 9 | Machines works fine, but blank/garbage display | | |
| 10 | Setting is not possible. The Microcontroller does not responding to push button pressing. | | |
| 11 | Overheating of Heating element/Transformer | | |
| 12 | Machine trips while pump starts working or abnormal sound of pump. | | |
| 13 | | Lid doesn’t open after | |
| | completing the | |
| | operation. | |
| 14 | Display shows fresh air always | Display shows fresh air | |
| | always | |
| | | |
| 15 | Fuse blown out when machine is switched on | | |
| | | |
| 16 | 24V Transformer Fuse blown out. | | |
| Fuse blown out when | | | |
| machine is switched on | | | |
| 24V Transformer Fuse | | | |
| blown out. | | | |